Options




Standard XE Head

 





The standard XE head is the basic head developed for all of the standard and the advanced modes for XE-series AFMs.

Specification

  • Z scan range: 12 µm
  • Resonant frequency of Z-scanner: 3 kHz
  • Laser type: LD (650 nm) or SLD  (830 nm), standard
  • Noise floor: 0.02 nm (typical), 0.05 nm (maximum)




25um XE Head

 





Extended scanning range of the 25 µm Z-scanner enables the measurement of high aspect ratio samples such as optical lenses and MEMS device. The head is fully compatible with all the existing modes and options.

Specification

  • Z scan range: 25 µm
  • Resonant frequency: 1.7kHz
  • Laser type: LD (650 nm) or SLD (830 nm)
  • Noise floor: 0.03 nm (typical), 0.05 nm (maximum)




XE Optical Head

 





The cantilever of the AFM can be used as a medium of light amplification when it is combined with NSOM or Raman spectroscopy, which enhances the optical response of the sample. In order to deliver maximized optical beam to and from the cantilever, the XE Optical Head provides wide optical accessibility from top, bottom, and side. It is also compatible with all the options of XE-series AFM.

Specification

  • Optical accessibility: top and side
  • Z scan range: 12 µm or 25 µm
  • Resonant frequency: 3 kHz (12 µm XE Head), 1.7 kHz (25 µm XE Head)
  • Laser type: LD (650 nm) or SLD (830 nm)
  • Noise floor: 0.03 nm (typical), 0.05 nm (maximum)




Hysitron Triboscope Adaptor Head

 





The adaptor head is designed to integrate Triboscope nanoindenter of Hysitron, Inc. with XE-series AFMs. The high feedback performance of the Z-scanner enables precise nanoindentation measurement.

Specification
  • Adaptor to combine with Triboscope nanoindenter of Hysitron, Inc. with the XE-Series
  • Z scan range: 12 µm or 25 µm
  • Resonant frequency: 3 kHz (12 µm XE Head), 1.7 kHz  (25 µm XE Head)
  • Vertical drift rate: < 1 nm/min




5 μm x 5 μm XY Scanner

 





For high resolution AFM/STM imaging, Park Systems offers the 5 µm XY-scanner.  This XY-scanner is only applicable to XE-70 and XE-100.

  • Scan range: 5 µm
  • Resonant frequency: > 850 Hz
  • Out-of-plane motion: < 1 nm
  • Resolution: 0.01 nm (open-loop)




50 μm x 50 μm XY Scanner

 





The 50 µm XY-scanner is the standard scanner for XE-70 and XE-100.  High precision scanning of the 50 µm XY-scanner provides stable imaging condition with no background curvature.

  • Scan range: 50 µm, (5 µm in low-voltage mode) 
  • Resonant frequency: > 850 Hz                 
  • Out-of-plane motion: < 1 nm
  • Resolution: 0.6 nm (closed-loop), 0.02 nm (open-loop)




100 μm x 100 μm XY Scanner

 





The 100 µm XY-scanner is developed to provide larger, yet precise, measurable area, which is indispensable for semiconductor and materials research. High precision, linearity, and orthogonal scanning performance of the scanner satisfies even the highest standard of industrial metrology.

  • Scan range: 100 µm, (10 µm in low-voltage mode)
  • Resonant frequency: > 450 Hz
  • Out-of-plane motion: < 2 nm
  • Resolution: 1.5 nm (closed-loop), 0.04 nm (open-loop)




100 μm x 100 μm XY Scanner for XE-150

 





The 100 μm XY-scanner allows wide scan range for materials and semiconductor research. Despite the large range, the scanner shows very low out-of-plane motion as well as high precision movement, linearity, and orthogonality.

  • Scan range: 100 µm
  • Resonant frequency: > 450 Hz
  • Out-of-plane motion: < 2 nm
  • Resolution: 1.5 nm (closed-loop), 0.04 nm (open-loop)




100 μm x 100 μm XY Scanner for XE-Bio

 





The 100 um XY-scanner of the XE-Bio provides wide observable range for biological samples. Based on the flexure scanner of the XE Technology, its design allows user convenient optical access to an inverted optical microscope for transparent samples.

  • Scan range: 100 µm
  • Resonant frequency: > 450 Hz
  • Out-of-plane motion: < 2 nm
  • Resolution: 1.5 nm (closed-loop), 0.04 nm (open-loop)




100 μm x 100 μm XY Scanner for NSOM/Raman


 





The 100 um XY-scanner for XE-NSOM/Raman is a large scale scanner with high precision, linearity, and orthogonal performance. The scanner allows for an optical path through the scanner for advanced optical measurements including Near-field Scanning Optical Microscopy (NSOM), Raman Spectroscopy, and Confocal Microscopy as well.

• Scan range: 100 μm
• Resonant frequency: > 450 Hz
• Out-of-plane motion: < 2 nm
• Resolution: 1.5 nm (closed-loop), 0.04 nm (open-loop)




Standard Probehand

 





The standard probehand is designed to be used in general operation. It features fast and easy cantilever loading of pre-mounted cantilevers and the cantilever bias function without wiring. The guided mounting of pre-mounted cantilevers also enables users to align the laser easily on the cantilever. The probehand supports all the standard and advanced modes except STM, SCM, and liquid imaging.

  • Cantilever mounting: Guided kinematic mounting
  • Cantilever bias range: -10 V - 10 V




Clip Type Probehand

 





Unmounted cantilevers can be used with the clip-type probehand. The probehand also has cantilever bias function without wiring, and supports all the standard and advanced modes except STM, SCM, and liquid imaging.

 • Unmounted cantilever can be used
 • Tip bias range: -10 V - 10 V
 • Tip bias function available for EFM and Conductive AFM
 • Supports all the standard and advanced modes except STM, SCM and in-liquid imaging






Liquid Probehand

 





The Liquid Probehand is designed for general liquid imaging in the contact mode as well as in the noncontact mode, and can form a closed liquid cell environment when used with the Universal liquid cell. Resistant to most of buffer solutions including acids, it is ideal for biological research.

  • Contact and Non-contact AFM imaging in liquid environment   
  • Allows closed-cell environment when used with Universal Liquid Cell   
  • Compatible with Standard chip carrier and Standard clip-type chip carrier   
  • Chemically resistive to acid/base conditions   
  • Compatible to biological samples 




SCM Probehand

 





Scanning capacitance microscopy measures the doping concentration of semiconductors by means of the change in the capacitive coupling between the cantilever and the sample. SCM probehand is made with dielectrics to minimize parasitic capacitive coupling.

  • Cantilever mounting: electrically conductive cantilevers on insulating ceramic carrier
  • Cantilever bias range: -10 V - 10 V
  • Designed to transfer RF signal to the cantilevers
  • Designed to minimize parasitic capacitance coupling





STM Probehand

 





A sharpened metal (Pt-Ir) tip instead of a cantilever is used for scanning tunneling microscopy (STM). The STM probehand is developed to hold the STM tip, and is shielded with ceramics to suppress possible electrical noise..

  • Tip mounting: Snapping of Pt-Ir wire
  • Tip bias range: -10 V - 10 V
  • Designed to minimize current noise




Universal Liquid Cell

 





The universal liquid cell was developed to meet diverse environmental control options with a single unit. The cell can be used as an open liquid cell as well as a closed liquid cell, while controlling the temperature. The cell includes three  outlets to perfuse liquid or gas to/from the cell, and two electrodes utilized as the reference and the counter electrode in electrochemistry.

Applicable configurations: open liquid cell, closed liquid cell, open liquid cell with heating/cooling, closed liquid cell with heating/cooling, electrochemistry cell, electrochemistry cell with heating/cooling

  • Open or closed cell with liquid/gas perfusion
  • Temperature control stage: 4 °C - 110 °C (in air), 4 °C - 70 °C (with liquid)
  • Chemically resistive to acid/base
  • Sample size: 20 mm in diameter / 5 mm in thickness
  • Electrochemistry upgrade
  • Applicable sample bias voltage: -10V - 10V




Electrochemistry Cell

 





  • Material: PCTFE, resistant against corrosive solutions
  • Number of electrodes: 4 electrodes
  • Applicable bias voltage: -10 V - 10 V
  • Capacity of liquid: 5 cm3
  • Sample size: 10 mm or 45 mm in diameter / 2 mm thickness




Open Liquid Cell

 





Open liquid cell is a cell which is used to accomodate liquid environment. The cell is resistant against corrosive solutions, and has clips to hold samples.

  • Material: PCTFE, resistant against corrosive solution
  • Capacity of liquid: 2.3 cm3
  • Sample size: 14 mm in diameter, or 20 mm × 35 mm / 4.5 mm thickness




Temperature Control Stage 1

 





Many polymers and ceramics show phase changes as the temperature goes through the critical temperature. The phenomena can be observed by using AFM with a heating/cooling stage option, which controls the temperature from 0 °C to 180 °C with 0.1 °C resolution.

  • Active Heating / Active Cooling
  • Temperature range: 0 °C - 180 °C
  • Temperature resolution: 0.1
  • Coolant circulation: Vibration-free water circulation
  • Sample size: 15 mm × 15 mm / 15 mm in thickness




Temperature Control Stage 2

 





Heating stage provides higher temperature range up to 250 °C.

  • Active Heating / Passive Cooling
  • Temperature range: ambient - 250 °C
  • Temperature resolution: 0.1 °C
  • Sample size: 15 mm × 15 mm / 15 mm in thickness




Temperature Control Stage 3

 





Heating stage provides higher temperature range up to 600 °C

  • Active Heating /  Passive Cooling
  • Temperature range: ambient - 600 °C
  • Temperature resolution: 0.1 °C
  • Sample size: 8 mm × 8 mm / 3 mm in thickness




Environmental Chamber

 





Environmental chamber provides controlled atmosphere for the samples sensitive to oxygen and/or water vapor. The atmosphere of the chamber is controlled by the organized functions of the humidifier, the dehumidifier combined with purging nitrogen gas into it.

  • Humidity control range: 2 % to 90 %
  • Gas inlet/outlet: 4 ports
  • Antechamber




Humidity Control System

 





  • Humidity control range: from 2 to 90%





Live Cell Chamber

 





The live cell chamber creates an ideal environment for cells, improving the life expectance during long measurement durations. The chamber controls temperature, humidity, and pH.

  • Temperature control: RT - 60 ¡ÆC (0.1 ¡ÆC stability)
  • Incorporated with gas mixer and humidity controller




Standard Acoustic Enclosure

 





▪ Environmentally sealed acoustic enclosure to block external acoustic and light noise
▪ Ergonomic design for user convenience
▪ Dimension: 510 × 655 × 720 mm (outer)
▪ Weight: 44.6 kg

For XE-70/100 only




Acoustic Foam Box

 





▪ Economic and highly effective solution for isolation from external acoustic and light noise
▪ Acoustic foam is used on the inside wall
▪ Dimension: 620 × 620 × 760 mm (outer)
▪ Weight: 7.4 kg

For XE-70/100 only




Integrated Acoustic Enclosure with Granite Table

 





Designed exclusively for the XE series system, the Integrated Acoustic Enclosure with Granite Table isolates the XE system from external acoustic and light noise for ultimate performance, and provides increased vibration isolation when used with Active Vibration Isolation Table.

▪ Includes a granite table for increased vibration isolation
▪ Best solution for sub-nanometer scale high resolution imaging
▪ Ergonomic design for a convenient access to the instrument
▪ Dimension: 820 × 920 × 1,345 mm (outer)

For XE 70/100/150 only




Acoustic Enclosure for Raman-AFM Integration

 





Environmentally sealed acoustic enclosure blocks external acoustic and light noise. Ergonomically designed for user convenience, the acoustic enclosure for Raman-AFM integration provides a beam path to an external spectrometer

▪ Dimension: 825 × 1000 × 730 mm (outer)
▪ Weight: 74.2 kg

 For XE-70/100/150 only




Acoustic Enclosure with Enhanced MOD and Side Opening Door

 





In this acoustic enclosure with steel frame and side opening door, enhanced MOD layers support both AFM and granite, giving the ultimate noise performance below 0.05 nm in the factory.

▪ Velocity Feedback: Electromagnetic transducers
▪ Frequency range: active 1 ~ 200 Hz, passive > 200 Hz
▪ System noise: less than 50 ng per root Hz from 0.1 to 200 Hz in all direction

For XE-150 only




Integrated Acoustic Enclosure for XE-Bio and XE-120

 





Designed exclusively for the XE-Bio and the XE-120, the Integrated Acoustic Enclosure for XE-Bio and XE-120 isolates the systems from external acoustic and light noise for ultimate performance, and provides higher degree of vibration isolation when used with Active Vibration Isolation Table.

▪ Includes a granite table
▪ Best solution for high resolution in-liquid imaging
▪ Ergonomic design for a convenient access to the instrument
▪ Dimension: 1,000 × 1,000 × 1,400 mm (outer)
▪ Weight: 646 kg

For XE-Bio/120 only




Active Q Control

 





During AFM imaging in Non-contact and tapping imaging, especially in liquid, the kinetic energy of the cantilever is decreased due to the interaction between the cantilever and the sample. By actively compensating the energy loss, Q control enhances the quality of topographic data. Active Q control circuit also provides arbitrary phase shift of the cantilever oscillation, by which the quality factor is increased or decreased.

  • Q enhancement ratio: 1 - 5   times (typical)
  • Q attenuation ratio: 10-2 - 1
  • Phase shift range: -180° - 180°




Signal Access Module

 





  • Enables access to various input/output signals for AFM
  • Scanner driving signal for the XY and Z scanners
  • Position signal for the XY and Z scanners
  • Cantilever deflection signals of the vertical/lateral direction
  • Bias signal for the sample and the cantilever
  • Driving signal for XE-AFM
  • Auxiliary input signal to the system

For the product datasheet of Signal Access Module, please click here.






External High Voltage Kit

 





The external high voltage kit provides an applied external bias option up to 2 kV.

 • Bias range: up to 2 kV
 • Works with Conductive AFM, EFM/SKPM/DC-EFM/PFM, and Nanolithography




Photon Counter

 





Photons generated by the laser excitation in XE-NSOM is delivered to the photon detector such as PMT or APD. The photon signal from the photon detector is transferred to the photon counter, whose counting is recorded as NSOM data with AFM data synchronously.

  • 2 input channels
  • Maximum counting per AFM pixel: 216
  • Minimum pulse width: 10 nsec
  • Time constant: 1 µsec - 224 µsec




Magnetic Field Generator

 





The magnetic field generator is used for applying external magnetic field to the sample. The field can be changed from -300 gauss to 300 gauss, and is parallel to the sample surface. The change in magnetic structure by the varying field can be observed by magnetic force microscopy (MFM).

  • Applying external magnetic field parallel to sample surface
  • Tunable magnetic field
  • Range: -300 - 300 gauss
  • Composed of pure iron core & two solenoid cells




Chip Carriers

 





  • Standard chip carrier
  • Ceramic chip carrier for SThM
  • Teflon coated chip carrier for EC-Cell
  • Ceramic chip carrier for SCM
  • Teflon coated chip carrier for Conductive AFM




Vacuum Chuck

 





For secure loading of samples, Park Systems provides a vacuum chuck. The chuck can manage 2 inch, 4 inch, and/or 6 inch wafers, and can be offered as a customized version.




Non-magnetic Sample Holder

 





The magnetic field from the standard sample holder may cause unintended results in delicate MFM measurement. To avoid this, non-magnetic sample holder is provided.




Atomic Force Microscope
Copyright © 2008 Park Systems Corp. All Rights Reserved.
Contact us | Site IndexRegister | Home